DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 72 of 72
03-2011 Type
5366/100
Agfa Company Confidential
MODULE : eboxfanmodule
Elabel : No
HC320000
: eboxfanmodule.eboxfan.connection
Type
Defect
Context
E320000
Action
SeeManual
Reason1
EBox fan module not connected
Cure1
Connect EBox fan module
Reason2
EBox fan module defect
Cure2
Replace
Part
MOD20
HC320020
: eboxfanmodule.eboxfan.shortcircuit
Type
Defect
Context
E320020
Action
SeeManual
Reason
EBox fan module defect
Cure2
Replace
Part
MOD20
MODULE : bothtraysmodule
Elabel : No
HC321001
: pickupmodule.alltrays.disabled
Type
Defect
Context
E321001
Action
SeeManual
Reason
All inputtrays have detected a mallfunction
Cure
Check other active HC-codes to get detailed info