DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 16 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC303012
: upperpickupmodule.elabel.present
Type
Failure
Context
W303012
Action
None
Reason1
Upper pickup elabel not connected
Cure1
Connect upper pickup elabel
Reason2
Upper pickup Elabel defect
Cure2
Replace
Part
MOD03
HC303013
: upperpickupmodule.elabel.config
Type
Failure
Context
W303013
Action
None
Reason
Upper pickup elabel defect
Cure
Replace
Part
MOD03
HC303014
: upperpickupmodule.connected
Type
Failure
Context
W303014
Action
None
Reason1
XS1
disconnected
Cure
Connect
XS1
Reason
Upper pickup module defect
Cure
Replace
Part
MOD03