Generic Safety Directions
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
Generic Safety Directions for HealthCare Imaging Products
Page 27 of 30
Sharp edges:
CAUTION:
Sharp edges inside the device: Cut or abrasion possible.
Be careful at maintenance and replacement of parts.
Cleaning optical elements:
CAUTION:
Image artifacts possible after cleaning optical elements.
When cleaning optical elements follow the service manual precisely.
Secured screws:
CAUTION:
Opening screws secured by red lacquer may misalign important device
adjustments:
Do not open screws that are secured by red lacquer.
Opening PCs and Workstations:
WARNING:
Electrical shock and damage to the equipment possible.
Only open the PC or workstation if explicitly stated in the service manual.
Unplug before opening.
Observe anti-static safety regulations.
Replacing fuses:
WARNING:
Replacing fuses by wrong type may lead to fire hazard.
Use only fuses of the exact value and characteristics stated in the service manual or on
the device.