![Advanced Instruments 3900 Service Manual Download Page 39](http://html.mh-extra.com/html/advanced-instruments/3900/3900_service-manual_2856019039.webp)
NOTE
A tube-loaded cassette must be
used to run the following test. Place
the cassette in the (left) input deck,
with the cassette identification
number facing you.
6. From the test menu, select
“Tube
Transport Test”
(see User Guide
Calibration section). Press START to begin
the cycling. Allow the test to run through
the entire cassette while adjusting for any
misalignments (using the sensor adjusting
screw).
7. Tighten the two slotted locking screws.
8. Place a cassette in the input deck and run
through a normal sample test cycle. Check
for any misalignments between the sample
tube and cooling well.
9. Adjust the cooling well position by
loosening the three slotted screws (C),
loosening the V-block bracket and V-block
screws (D), and manually moving the well
into a position that minimizes/eliminates
any misalignment with the operating head.
NOTE
If misalignment cannot be
adjusted by moving the well, the
position of the head will require
readjustment, and the entire
process from the insert mode
position adjustment must be
repeated.
39
Insert Mode Position
1. Enter the head rotation test from the test
menu and select <YES> for
“Insert Test”
.
Check the alignment of the operating head
in the insert well. Note the direction and
the amount of misalignment, if any.
2. Remove the back cover from the
instrument.
3. Manually rotate the head to probe wipe
position and slightly loosen the cap screw
(A) on the anti-rotation block.
4. Manually rotate the operating head over
the cooling well, using the position over
the well as a reference point.
5. Move the position of the head by noted
amount of misalignment. Retighten the cap
screw.
6. Readjust the operating head (position 1) as
described below.
7. Retest using the head rotation test (see
User Guide Calibration section) and
readjust, if necessary.
Operating Head (position 1)
1. Loosen the two slotted locking screws (B)
located above the sensor adjustment screw
(i.e., on the right side of the head rotation
cap).
2. From the test menu, select
“Head
Rotation Test”
(see User Guide
Calibration section). Press START to
begin cycling.
3. Note the direction and amount of the
misalignment of the probe from the center
of the tube sensor during the cycling.
4. Turn the sensor adjustment screw (A)
counterclockwise to move the head
position to the right, or clockwise to move
the head to the left.
5. Adjust until the sample probe is centered
with the tube sensor.
The Advanced
®
Model 3900 Service Manual
C
B
A
D
Summary of Contents for 3900
Page 1: ...3905SM Rev14 050415 The Advanced Osmometer Model 3900 Service Manual ...
Page 2: ......
Page 5: ...1 Introduction ...
Page 6: ......
Page 13: ...2 General Overview ...
Page 14: ......
Page 15: ...The Advanced Model 3900 Service Manual 15 Circuit Diagram ...
Page 20: ...20 The Advanced Model 3900 Service Manual Notes ...
Page 21: ...3 Maintenance ...
Page 22: ......
Page 25: ...4 Troubleshooting ...
Page 26: ......
Page 36: ...36 The Advanced Model 3900 Service Manual Notes ...
Page 37: ...5 Adjustments ...
Page 38: ......
Page 41: ...6 Replacement Instructions ...
Page 42: ......
Page 44: ......
Page 46: ......
Page 50: ......
Page 54: ......
Page 68: ......
Page 74: ......
Page 84: ......
Page 100: ......
Page 104: ......
Page 113: ...4CP450 Rev1 4C35 3905 Service Manual Page 3 of 3 E J12 J3 J11 J13 ...
Page 114: ......
Page 118: ......
Page 130: ......
Page 138: ......
Page 148: ......
Page 152: ......
Page 154: ...4CP446 Rev0 4C35 3905 Service Manual Page 2 of 2 C D E ...
Page 159: ...7 Schematics ...
Page 160: ......
Page 162: ......
Page 169: ...Appendix A Symbol Definitions ...
Page 170: ......
Page 173: ...Appendix B Product Disposal and Recycling ...
Page 174: ......
Page 176: ...The Advanced Model 3900 Service Manual Notes ...
Page 177: ...Appendix C Service Log ...
Page 178: ......
Page 180: ...Symptoms or Problems Repairs Replacements or Changes Serviced by ...
Page 181: ...Index ...
Page 182: ......