18
The fluid level probes are interfaced to a
peripheral interface adapter through discrete
logic. The peripheral interface adapter
interfaces with the microprocessor. There are
three LEDs located on the motor control
board which indicate the fluid level status.
• Yellow on indicates fluid level low.
• Yellow/ green on indicates fluid level half.
• Red/green/ yellow on indicates fluid level
full.
The status of the fluid level can also be
checked by using the “reservoir status”
function in the setup menu.
Serial Communication Port
Serial RS-232 Port Interface:
An Asynchronous Communications Interface
Adapter and RS-232 line drivers/receivers
provide the microprocessor with a serial port
interface that supports both hardware and
software handshaking.
The RS-232 port has the following pin
assignments:
Signal
DB-9 Pin
Carrier Detect
1
Receive Data
2
Transmit Data
3
Data Terminal Ready
4
Signal Ground
5
Data Set Ready
6
Request to Send
7
Clear to Send
8
Head Rotation System
The disk sensor circuit controls the position
of the sample head at the cassette tube
retrieval/placement position. The controlling
system consists of an optical switch
(adjustable) and a slotted disk. The optical
switch is interfaced to the microprocessor
through a peripheral interface adapter.
The head rotation travel is controlled by two
mechanical limit switches; one is located on the
right side and one on the left side. The limit
switches are interfaced to the microprocessor
through a peripheral interface adapter.
Barcode Port:
A D-type, 15-pin barcode port is provided in
the back of the instrument for connecting and
providing power to such a device. For proper
operation, the barcode port requires a 1200
bps TTL signal providing asynchronous serial
data containing 1 start bit, 8 data bits, 1 stop
bit, and no parity.
A suitable barcode scanner is available from
Advanced Instruments. To interface with the
instrument, the barcode scanner must be
programmed as follows, referring to the
scanner user’s guide.
1200 bps
CR suffix
disable beep after good decode
triggerless trigger mode (optional)
The Advanced
®
Model 3900 Service Manual
+5V DC
1
to reader
receive data
10
from reader
gnd/earth
9
common
Signal
Pin
Direction
Barcode Port Connections
Summary of Contents for 3900
Page 1: ...3905SM Rev14 050415 The Advanced Osmometer Model 3900 Service Manual ...
Page 2: ......
Page 5: ...1 Introduction ...
Page 6: ......
Page 13: ...2 General Overview ...
Page 14: ......
Page 15: ...The Advanced Model 3900 Service Manual 15 Circuit Diagram ...
Page 20: ...20 The Advanced Model 3900 Service Manual Notes ...
Page 21: ...3 Maintenance ...
Page 22: ......
Page 25: ...4 Troubleshooting ...
Page 26: ......
Page 36: ...36 The Advanced Model 3900 Service Manual Notes ...
Page 37: ...5 Adjustments ...
Page 38: ......
Page 41: ...6 Replacement Instructions ...
Page 42: ......
Page 44: ......
Page 46: ......
Page 50: ......
Page 54: ......
Page 68: ......
Page 74: ......
Page 84: ......
Page 100: ......
Page 104: ......
Page 113: ...4CP450 Rev1 4C35 3905 Service Manual Page 3 of 3 E J12 J3 J11 J13 ...
Page 114: ......
Page 118: ......
Page 130: ......
Page 138: ......
Page 148: ......
Page 152: ......
Page 154: ...4CP446 Rev0 4C35 3905 Service Manual Page 2 of 2 C D E ...
Page 159: ...7 Schematics ...
Page 160: ......
Page 162: ......
Page 169: ...Appendix A Symbol Definitions ...
Page 170: ......
Page 173: ...Appendix B Product Disposal and Recycling ...
Page 174: ......
Page 176: ...The Advanced Model 3900 Service Manual Notes ...
Page 177: ...Appendix C Service Log ...
Page 178: ......
Page 180: ...Symptoms or Problems Repairs Replacements or Changes Serviced by ...
Page 181: ...Index ...
Page 182: ......