34
The Advanced
®
Model 3900 Service Manual
Symptom
Check
Suggested Action
“
Sample Pre-freeze
”
1. Were the sample probe and tube
clean? Are the probe wipers dirty?
2. Is the probe configuration correct?
3. Check the stir/freeze wire
alignment and adjustment.
4. Is the sample frozen?
5. Does the clapper and/or the stir
wire coil show excessive wear?
1. Replace the probe wipers and test
another aliquot of that sample in a
clean sample tube. Check/correct
the probe bin numbers.
2. Correct the stir/freeze wire
alignment and adjustment, as
necessary.
3. Try to minimize the effects of
materials that tend to cause
premature crystallization.
4. False prefreeze errors may be
reported if the instrument cannot
maintain a cooling sample rate of
greater than 20 mOsm per 7
seconds.
5. Contact Advanced Instruments
for Hot-Line
®
Technical Service,
and possible replacement of the
yoke, clapper, and coil by a
qualified service technician.
“
Sample Probe Failure
”
Switch the instrument off, then on.
Check the sample probe by running
the A/D Test.
If the error message does not persist
and other messages are not
displayed, ignore this message.
Replace the sample probe.
“
Sample Probe Open?
”
Switch the instrument off, then on.
Check the sample probe by running
the A/D Test.
If the error message does not persist
and other messages are not
displayed, ignore this message.
Replace the sample probe.
“
Sample Tube Error
”
Check for a sample tube in the wrong
place, possibly in the operating head
or freezing chamber.
Review procedure and retry the test.
Remove the misplaced tube, check
the probe and stir-freeze wire
alignment, and retry the test.
If the error message persists,
contact Advanced Instruments
for Hot-Line
®
Technical Service.
“
Serial Loop Failed
”
1. Is serial loop-back connector
attached?
2. Did you wire loop-back connector
correctly?
3. Try lower baud rate.
1. Attach connector.
2. See User Guide, Chapter 5.
3. Change baud rate.
Summary of Contents for 3900
Page 1: ...3905SM Rev14 050415 The Advanced Osmometer Model 3900 Service Manual ...
Page 2: ......
Page 5: ...1 Introduction ...
Page 6: ......
Page 13: ...2 General Overview ...
Page 14: ......
Page 15: ...The Advanced Model 3900 Service Manual 15 Circuit Diagram ...
Page 20: ...20 The Advanced Model 3900 Service Manual Notes ...
Page 21: ...3 Maintenance ...
Page 22: ......
Page 25: ...4 Troubleshooting ...
Page 26: ......
Page 36: ...36 The Advanced Model 3900 Service Manual Notes ...
Page 37: ...5 Adjustments ...
Page 38: ......
Page 41: ...6 Replacement Instructions ...
Page 42: ......
Page 44: ......
Page 46: ......
Page 50: ......
Page 54: ......
Page 68: ......
Page 74: ......
Page 84: ......
Page 100: ......
Page 104: ......
Page 113: ...4CP450 Rev1 4C35 3905 Service Manual Page 3 of 3 E J12 J3 J11 J13 ...
Page 114: ......
Page 118: ......
Page 130: ......
Page 138: ......
Page 148: ......
Page 152: ......
Page 154: ...4CP446 Rev0 4C35 3905 Service Manual Page 2 of 2 C D E ...
Page 159: ...7 Schematics ...
Page 160: ......
Page 162: ......
Page 169: ...Appendix A Symbol Definitions ...
Page 170: ......
Page 173: ...Appendix B Product Disposal and Recycling ...
Page 174: ......
Page 176: ...The Advanced Model 3900 Service Manual Notes ...
Page 177: ...Appendix C Service Log ...
Page 178: ......
Page 180: ...Symptoms or Problems Repairs Replacements or Changes Serviced by ...
Page 181: ...Index ...
Page 182: ......