For 4C3 Cryoscopes & 3900 Osmometers:
1. Reconnect the fan and reinstall the
instrument back cover.
2. Reconnect the power cord and turn the
instrument on.
3. Make sure the Supervisor/Operator
keyswitch is in the
Supervisor
position
while using the
TEST
and
CALIB
menus
(see User Guide, Chaper 2).
4. When
“
Cryoscope
(or
Osmometer
)
Ready
”
is displayed, press the
TEST
soft
key. The display will change to
“
Select
Test Item
”
.
5. At
“
Select Test Item
”
, repeatedly press
the
>
soft key until the display reads
“
Probe Bin Test
”
.
6. At
“
Probe Bin Test
”
, press
START
. The
display will change to
“
[Start] Ready?
”
.
7. At
“
[Start] Ready?
”
, pipet or measure a
sample of probe bin-setting fluid into a
clean sample tube, place the tube in the
first position in a cassette.
8. Place the cassette in the left (input) deck
and press the
START
soft key. The
instrument will run a special freezing-
point test to determine the sample probe
resistance and bin number.
9. Record the sample probe resistance and
bin number in the Service Log at the end
of the user’s guide; then press the
STOP
soft key. The display will change to
“
Probe Bin Test
”
.
10. Press
SETUP
to enter the SETUP menu,
then repeatedly press the
>
soft key until
the display reads
“
Set Sample Bin #
”
.
11. At
“
Set Sample Bin #
”
, press the
START
soft key. The display will report
the current sample bin number setting.
12. If the current sample bin number setting
is not the same as that just determined,
enter the correct bin number by means of
the numeric keypad; then press the
ENTER
soft key. The display will change
to
“
Set Sample Bin #
”
; then press the
STOP
soft key to return to
“
Cryoscope
(or
Osmometer
)
Ready
”
. If the current
sample probe bin setting is the same as
that just determined, simply press the
STOP
soft key twice to return to
“
Cryoscope
(or
Osmometer
)
Ready
”
.
Press the
CALIB
soft key and calibrate
the instrument as recommended in the
User Guide, Chaper 4.
NOTE:
Recalibration using the
CALIB
procedure is required when the
sample probe is replaced; trying
to use the automatic calibration
cassette when replacing the sample
probe may cause a
“
Calibrator
Out of Range
”
error message.
4CP700 Rev5
(4C35/3905 Service Manual)
Page 5 of 5
Summary of Contents for 3900
Page 1: ...3905SM Rev14 050415 The Advanced Osmometer Model 3900 Service Manual ...
Page 2: ......
Page 5: ...1 Introduction ...
Page 6: ......
Page 13: ...2 General Overview ...
Page 14: ......
Page 15: ...The Advanced Model 3900 Service Manual 15 Circuit Diagram ...
Page 20: ...20 The Advanced Model 3900 Service Manual Notes ...
Page 21: ...3 Maintenance ...
Page 22: ......
Page 25: ...4 Troubleshooting ...
Page 26: ......
Page 36: ...36 The Advanced Model 3900 Service Manual Notes ...
Page 37: ...5 Adjustments ...
Page 38: ......
Page 41: ...6 Replacement Instructions ...
Page 42: ......
Page 44: ......
Page 46: ......
Page 50: ......
Page 54: ......
Page 68: ......
Page 74: ......
Page 84: ......
Page 100: ......
Page 104: ......
Page 113: ...4CP450 Rev1 4C35 3905 Service Manual Page 3 of 3 E J12 J3 J11 J13 ...
Page 114: ......
Page 118: ......
Page 130: ......
Page 138: ......
Page 148: ......
Page 152: ......
Page 154: ...4CP446 Rev0 4C35 3905 Service Manual Page 2 of 2 C D E ...
Page 159: ...7 Schematics ...
Page 160: ......
Page 162: ......
Page 169: ...Appendix A Symbol Definitions ...
Page 170: ......
Page 173: ...Appendix B Product Disposal and Recycling ...
Page 174: ......
Page 176: ...The Advanced Model 3900 Service Manual Notes ...
Page 177: ...Appendix C Service Log ...
Page 178: ......
Page 180: ...Symptoms or Problems Repairs Replacements or Changes Serviced by ...
Page 181: ...Index ...
Page 182: ......