68
Human NIBP Nano
Owner’s Guide
Electromagnectic Immunity – Guidance and manufacturer’s declaration
The Human NIBP Nano Interface is intended for use in the electromagnetic
environment specifi ed below. The customer or the user of the Human NIBP Nano
Interface should assure that it is used in such an environment.
Immunity Test
IEC 60601 test
level
Compliance level
Electromagnetic
environment – guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile.
If fl oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5%
Ut
(>95% dip
in
Ut
for 0.5 cycle)
40%
Ut
(60% dip in
Ut
) for 5 cycles
70%
Ut
(30% dip in
Ut
for 25 cycles
<5%
Ut
(>95% dip
in
Ut
) for 5 sec
< 5%
Ut
(>95% dip
in
Ut
for 0.5 cycle)
40%
Ut
(60% dip
in
Ut
) for 5 cycles
70%
Ut
(30% dip
in
Ut
for 25 cycles
<5%
Ut
(>95% dip
in
Ut
) for 5 sec
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the Human
NIBP Nano Interface
requires continued
operation during power
mains interruptions, it
is recommended that
the Human NIBP Nano
Interface be powered
from an uninterruptible
power supply or a
battery.
Note:
Ut
is the a.c. mains voltage prior to application of the test level.
Power frequency
(50/60 Hz)
magnetic fi eld
30 A/m
30 A/m
Power frequency
magnetic fi elds
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.