33
Human NIBP Nano
Owner’s Guide
•
Finger cuff switching is also disabled – the pressure waveforms
generated are applied at the C1 air hose outlet on the Wrist Unit.
•
The Test mode button has changed to indicate that Test mode is
on.
Available Test modes
1. Steady pressure –
This generates a flat, constant-pressure wave.
Available options are 50, 100, 150, 200, 250 and 300 mmHg.
•
In this test mode it takes about 20 seconds for the Human NIBP
Nano to build up to the correct mode and pressure.
Figure 2–18
Human NIBP HCU and
test dialog (right) showing
available options, and
the Human NIBP Settings
dialog (left) showing
AutoCal and fi nger cuff
switching options are
disabled in Test mode..