34
Human NIBP Nano
Owner’s Guide
2. Square wave –
This generates a repeating square pressure wave.
Available baseline pressure options are 50, 100, 150, 200 and 250 mmHg,
and frequency options are 1, 4 and 10 Hz. Note how, in Square wave mode,
the Human NIBP Nano Interface uses a fixed off set of ± 50 mmHg that
cannot be changed.
•
In Figure 2–20 a pressure off set of 100 mmHg at 1 Hz was selected.
The resultant square wave is 150 mmHg for 0.5 s, followed by 50
mmHg for 0.5 s, and so on.
Figure 2–19
Human NIBP Test Mode
dialog showing the
available Steady pressure
options (top) and the
generation of a steady
pressure of 100 mmHg
(bottom).
Figure 2–20
Human NIBP Test Mode
dialog showing the
Square wave options
(top) and the generation
of a repeating square
wave with a baseline