12
Human NIBP Nano
Owner’s Guide
•
To maintain operator- and subject-safety, peripheral equipment
that is connected to one of the Human NIBP Nano system
components should comply with IEC60601-1 and other relevant
safety standards.
•
The operator should never touch both the patient and the Human
NIBP Nano system at the same time.
•
For measurements longer than 1 hour, the use of finger switching
between two attached finger cuff s is strongly recommended since
a prolonged measurement on one finger can be unpleasant for the
subject.
•
The ADInstruments Human NIBP soft ware is designed to support
the connection of a single Human NIBP Nano unit at any one time.
•
Diseases or environmental conditions that result in reduced finger
arterial compliance (diabetes, Raynaud disease, cold fingers),
which causes reduced volumetric pulsations in the fingers, may
lead to an inability to measure blood pressure or derived variables.
This will be detected by the device which will report an error.
•
All cables should be routed in such a way that strangulation cannot
occur.
•
The Human NIBP Nano system is not defibrillation proof.
•
The user must take care to not use the system if any components
are damaged and also ensure that the Nano core wrist unit is never
submerged.
•
At high ambient temperatures, the temperature of the Nano Core
wrist unit can go up to 59 degrees Celsius.