Human NIBP Nano
Owner’s Guide
Instrument Information
Cuff pressure:
Max. 350 mmHg
Height sensing:
Range ± 100 mmHg
Blood pressure accuracy:
1% of full scale (max. 3 mmHg)
Zeroing
automatic
Instrument Accuracy
Height correction:
2 mmHg
Manual
zeroing
Heart rate:
(Rate [bpm] / 60)%, i.e., at 60 bpm, accuracy is ± 1%
Interbeat interval:
10ms (max, non-accumulating)
Applied Parts
•
Wrist Unit (plastic housing and velcro strap)
•
Cuff / Plethysmograph
•
Height Correction Unit
The EMC conformity of the Human NIBP Nano Interface includes the
use of the following external cables, transducers and accessories:
Cable/accessory
Description
Wrist Unit cable assembly,
including Height Correction Unit
and fi nger cuff connected to the
Wrist Unit
Standard length of Wrist Unit cable assembly: 275 cm.
Standard length of HCU cable: 30 cm.
The Human NIBP Nano Interface should not be used adjacent to or stacked with
other equipment and that if adjacent of stacked use is necessary, the Human
NIBP Nano Interface should be observed to verify normal operation in the
configuration in which it will be used.