35
Human NIBP Nano
Owner’s Guide
•
Therefore, if your pressure off set is 250 mmHg, what you see is 200
mmHg, then 300 mmHg. Alternatively, if your pressure off set is 50
mmHg, what you see is 100 mmHg, then 0 mmHg.
•
In this test mode it takes about 20 seconds for the Human NIBP
Nano to build up to the correct mode and pressure (you should feel
the finger cuff pulsating erratically as it starts up).
3. Simulate pressure –
the simulated pressure test mode does not
inflate the finger cuff s. It creates a continuous square waveform in the
Finger Pressure and HCU Pressure channels (see Figure 2–21 on page
35):
•
In the Finger Pressure channel, it generates a simulated pressure
wave of 2 second duration at 200 mmHg, followed by 3 seconds at
0 mmHg.
•
In the HCU Pressure channel, it generates a simulated pressure
wave of 2 second duration at 100 mmHg, followed by 3 seconds at
-100 mmHg.
Using Test Mode
In Steady pressure and Square wave modes, the pressure waveforms
generated by the Human NIBP Nano are applied at the C1 air hose outlet
of the Wrist Unit.
Figure 2–21
The Test Mode dialog
with Simulate pressure
selected.