To confirm that LabChart has found the Human NIBP Nano Interface:
•
The device name should appear at the top of the new document.
•
The Input Amplifier dialog should be disabled for all 11 LabChart
channels because data is streamed directly from the Human NIBP
Nano Interface. There is no need to change the settings, range or
display (see the LabChart Help for more details).
•
The
Start
button should be enabled and ready to start recording.
It is necessary to set up the hardware before proceeding to make a
finger blood pressure measurement. If you are already familiar with the
hardware, please proceed to Starting up the soft ware on page 23.
The Human NIBP Wrist Unit
The Wrist Unit (Figure 2–4) contains an air pressure control valve, a
pressure transducer, electronics for the infrared photoplethysmograph in
the finger cuff , and electronics for the Height Correction Unit (HCU).
The Wrist Unit is connected to the Human NIBP Nano Interface with a
single cable.
Figure 2–4
Human NIBP Nano Wrist
Unit with two fi nger cuffs
attached.
Note: the HCU (grey) is
also shown
Human NIBP
Nano Wrist Unit
[FMS910804]
Finger cuff
[MLT382/X]
C1
HCU
C2
Cuff cable
connector
Air hose
Height Correction
Unit (HCU)
[FMS903903]
Cable assembly