CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 110 of 126
© 2019 Abbott– Document No PL-0047, Rev 11 (December 2019)
8) Pumping operation may now be re-established via the backup Console by increasing
the RPM to about 1000 RPM. Gradually increase RPM while unclamping the Pump
outflow tubing. Continue to gradually increase the RPM to achieve the desired flow.
Pumping operation should be returned to the last operating condition of the Pump. Be
sure to reset options and alarms to match those selected before hardware exchange.
Figure 63: Emergency Switch to Backup System
If the Pump has been OFF for more than five minutes without adequate anticoagulation, or
if there has been a Motor Overheating condition, it will be necessary to replace the Pump
and circuit, including Cannulas.
NOTE:
To be available for emergency use, the backup Console, Motor and Flow Probe
must be transported with and in proximity of the main Console and Motor with the main
power switch OFF. The battery for the backup Console must be periodically assessed
according to