CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 83 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
The sub-folders are labeled “
xx-xx-xx-xx-xx-xx_JJJJ-MM-DD_hh-mm-ss
” where “
xx-xx-
xx-xx-xx-xx
” represents the Monitor hardware ID and “
JJJJ-MM-DD_hh-mm-ss
”
indicates the copy date and time (e.g. 00-00-10-74-61-3f_2010-01-25_14-25-36).
The logger data files are stored within these subfolders. The logger data file names are
labeled “
yy-yy-yy-yy-yy-yy_zzzz-zzzz-zzzz_aaaa.log
” where “
yy-yy-yy-yy-yy-yy
”
represents the Console hardware ID, “
zzzz-zzzz-zzzz
” represents the Console firmware ID
and “
aaaa
” indicates the logger session (e.g. 00-00-10-61-2f-9f_0021-0000-
003a_001e.log).
7.24 Monitor Connection
The Console may be used with or without the Monitor. It is recommended that the Monitor
be used whenever the patient is not being transported and is in a stationary mode. To use
the Monitor, connect the cable from the Monitor to the rear panel of the Console, and
mount the Monitor to the mounting bracket on the Cart. The Monitor is automatically
powered ON as soon as it is connected to the Console. The Monitor displays basic
operational data and System status updates when the System is operated in either the
biventricular or univentricular mode of operation.
The Monitor is designed to display information derived from one or two Consoles. It is
possible, therefore, to view data when the System is operated in the univentricular
configuration (LVAS or RVAS) or when the System is used in the BiVAS configuration
(LVAS + RVAS). To do so, the user must provide input to the Console, either during set-up
or at any time thereafter, by designating whether the Console is being used as an LVAD or
an RVAD. This is accomplished by accessing the
MENU
Options on the Console and
selecting
Application Mode
. Use the
UP
and
DOWN
arrows to select either LVAD or
RVAD. The information on the Monitor will be displayed in
Red
when a Console is
designated for use as an
LVAD
, in
Blue
if designated for use as an
RVAD.
CAUTION
The Monitor may be operated when the 2
nd
Generation CentriMag Console is
running on AC power or in combination with the Console Version 2 on battery.
Operating the Monitor while the Console runs on battery shortens the battery
runtime. Refer to the individual RVAD or LVAD Console display when the
Monitor is not in use.