CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 75 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
7.18 Setting the Console Speed Step Resolution
Pressure Display
Speed Step Resolution
Flow Range
Flow Limit Sensitivity
Flow Recorder Speed*
Set RPM
Exit
Menu
Active
Inactive
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Menu
Menu
Menu
Menu
Menu
Menu
Normal
(100 rpm)
Fine
(50 rpm)
Set RPM
Menu
R
Normal
Sensitive
Set RPM
Menu
Next
Normal
(0 – 10 lpm)
Reduced
(0 – 3 lpm)
Slow
(2 min)
Fast
(20 sec)
Set RPM
Menu
L
Select
Enter
Return
* Monitor only
Extended Menu
Set RPM
Menu
[ |Confirm]
[Execute|Cancel]
Set RPM
Menu
[ |Confirm]
[Start|Cancel]
Copy Data*
Management*
Support Type
Language
Data Logger*
Figure 46: Monitor MENU Structure
Speed Step Resolution
The System may be used to treat patients with a wide range of body sizes and conditions.
The
SPEED STEP RESOLUTION
function is provided to allow the user to select smaller
incremental changes in speed (50 RPM increments) for small individuals, and larger (100
RPM) changes for the larger patients. The default setting is the 100 RPM option. To
access the
SPEED STEP RESOLUTION
function, depress the
MENU
keypad and scroll
down to the
EXTENDED MENU
option, then depress the Menu key until the SPEED
STEP RESOLUTION option appears. Choose between the
STEP 50
(50 RPM) and
STEP
100
(100 RPM) options using the
UP
and
DOWN
arrows. Once selected, this will dictate
the speed that the Pump will increase with each depression of the
UP
or
DOWN
ARROW
keypads when using the
SET RPM
option.