CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 107 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
Note: If there are any issues while connecting the CentriMag Motor to the CentriMag
Console, inspect the Motor port on the rear of the Console for damage, particularly for any
broken pins that may be lodged inside.
2) Visually inspect the entire length of the CentriMag Motor cable, including both bend
reliefs (see Figures 57 and 58), for any damage such as separations of the bend reliefs,
deformations, kinks, or cuts. These types of damage indicate wear and tear or
previous rough handling of the cable, which has the potential to result in internal wire
damage. Cable thickness and shape should be uniform throughout the length of the cable.
Figure 57: Cable Bend Relief at the Console Connector
Figure 58: Cable Bend Relief at the Motor
Figure 59: Cable Damage at the Motor Bend Relief