Table 6: Features of the Console, Monitor, and Flow Probes
FEATURE
NAME
DESCRIPTION
LOCATION
Flow Probes Labels
IPX4
1/4“ x 3/32“
On the cover of the
Flow Probes
Defibrillation-
proof Type CF
Equipment
Equipment type
On the label of the
Flow Probe’s
cover
General
Warning Sign
Connect this external flow sensor to
the back panel of the Console only.
On the label of the
Flow Probe’s
cover
IPX4
Ingress
Protection
Rating
Protection against splashing water
On the label of the
Flow Probe’s
cover
6.7.4 Alarm and Alert Conditions
Audible and visual alarm/alert conditions warn the operator to conditions that may interrupt
patient support or damage the Pump or the Console. If an alarm/alert condition occurs, the
audible alarm/alert sounds, and an alarm/alert message indicating the cause(s) of the
alarm/alert appears on the display. Depressing the ALARM ACKNOWLEDGE keypad
mutes the audible alarm. The alarm/alert message will be continuously displayed on the
top two lines of
Row 2
of
the Console’s display as long as the alarm/alert condition exists.
In the event of an alert or alarm condition (see
or
alarms and alerts), visual message and audio indicators activate. The Console continues
Pump operation during an alert or
MOTOR ALARM
condition, and stops the Pump during
an alarm or
MOTOR STOPPED
condition. Both the visual and audio alert indicators are
active, even if the problem has resolved, until the alert/alarm condition is acknowledged by
pressing the ALARM ACKNOWLEDGE keypad. The user must acknowledge the
alert/alarm to silence the audio indicator and to determine if the alert/alarm condition has
resolved or is unresolved.
•
If the alert/alarm condition has not been resolved, pressing the ALARM
ACKNOWLEDGE keypad temporarily mutes the audio alert indication if the
alarm/alert condition is silenceable. The alarm/alert message will still be displayed
on the screen.
• If the alert/alarm condition has been resolved, pressing the ALARM
ACKNOWLEDGE keypad mutes the audio alert indication and removes the visual
alarm/alert message.