CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 93 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
message is displayed, but the Pump does NOT stop. The operator can mute the audible
alert by depressing the ACKNOWLEDGE keypad, which silences the audio advisory, but
will not remove the visual message.
The exception to audio alarm silencing is the
BATTERY BELOW MINIMUM
alert while
running on battery. This alert cannot be silenced until the Console is plugged into AC
power. The
BATTERY BELOW MINIMUM
alert will not stop the Pump until the battery is
fully discharged at which point the Console will power off. If this alert occurs, the user must
immediately plug the System into AC power, or change to a backup System.
If an alert condition persists for more than 60 seconds after the alert has been
acknowledged, the audio advisory reactivates and continues until acknowledgement
except for
BATTERY MAINTENANCE REQUIRED
(only requires acknowledgment once)
and
ON BATTERY
/
LOW BATTERY
(audio reactivates every 15 / 10 minutes or until
reconnected to AC power). The recommended action by the operator during an Alert
Condition is to take action to resolve the specific fault condition.
The visual display of the alert condition will be automatically removed provided: a) the
condition was acknowledged by depressing the Alarm Acknowledge button, and b) the
condition was resolved. If the visual indication of the alert condition remains, the user
either failed to acknowledge the condition by depressing the Alarm Acknowledge button,
the alert condition has resolved and reoccurred, or the condition has not been successfully
resolved.
Note: Should the alert condition be successfully resolved but the visual indication persists,
the user may clear the visual display by depressing the Alarm Acknowledge button.
Persistence of the visual display after depressing the Alarm Acknowledge button indicates
the alert condition has not been resolved or has reoccurred. If the alert condition cannot
be resolved the user should refer to
to determine the appropriate
operator response for the specific alert which has occurred.
8.3.3 Response to System Alarms or Alerts
When a System alarm or alert condition exists, an alarm tone sounds and a text message
appears on the display of the Console. Most alarms and alerts require some action on the
user’s part to correct the cause. The following table can be used to determine how to
correct an alarm or alert condition.
Table 13: Console Alarms & Alerts
ID
Alarm/
Alert
Text Message
System Status & Operator Response
S1
Alarm
POWER ON TEST FAIL
The Pump will not start.
An audible alarm will sound, which cannot be muted.
Switch the Console OFF and ON again. If the alarm re-appears,
switch to the backup Console, Motor and Flow Probe, record
the alarm message and contact your local Abbott Medical
representative.
S2
Alarm
SYSTEM FAULT
(Run-Time System Failure)
The Pump will stop.
An audible alarm will sound, which cannot be muted.
Clamp the return tubing and switch to the backup Console,
Motor and Flow Probe according the procedure described in
Section 10.1. Resume support. Record the alarm message and
contact your local Abbott Medical representative.
M1
Alarm
MOTOR STOPPED
The Pump will stop.
An audible alarm will sound, which can be muted for 60
seconds.
Clamp the return tubing and switch to the backup Console,