Table 13: Console Alarms & Alerts
ID
Alarm/
Alert
Text Message
System Status & Operator Response
Motor and Flow Probe according the procedure described in
Section 10.1. Resume support. Record the alarm message and
contact your local Abbott Medical representative.
M2
Alarm
MOTOR DISCONNECTED
The Pump will stop.
An audible alarm will sound, which can be muted for 60
seconds.
During setup of the System:
Press the alarm acknowledge button and check that the Motor
connector is fully inserted into the back of the Console.
During support:
Press the alarm acknowledge button and check that the
connector of the Motor is fully inserted into the back of the
Console. Resume support. If the visual alarm message does not
disappear, clamp the return tubing and switch to the backup
Console, Motor and Flow Probe according the procedure
described in Section 10.1. Resume support. Record the alarm
message and contact your local Abbott Medical representative.
M3
Alarm
PUMP NOT INSERTED
System will not start.
An audible alarm will sound, which can be muted for 60
seconds.
Press the alarm acknowledge button. Insert or re-insert the
Pump and secure it with the locking screw.
If the alarm repeats, switch to backup Console, Motor and
Flow Probe according the procedure described in Section 10.1.
M4
Alarm
MOTOR ALARM
An audible alarm will sound and the System will continue to
operate.
Press the alarm acknowledge button, if the visual alarm
message does not disappear, clamp the return tubing, stop the
Pump and switch to the backup Console, Motor and Flow
Probe according the procedure described in Section 10.1.
Resume support. Record the alarm message and contact your
local Abbott Medical representative.
M5
Alert
SET PUMP SPEED NOT
REACHED
Check the Pump flow: If Pump flow is satisfactory; reduce set
speed while insuring that flow is maintained.
Press the alarm acknowledge button. If the alert repeats,
clamp the return tubing, stop the Pump and switch to the
backup Console, Motor and Flow Probe according the
procedure described in Section 10.1. Resume support. Record
the alert message and contact your Abbott Medical
representative.
If the Pump flow is not satisfactory, clamp the return tubing
and switch to the backup Console, Motor and Flow Probe
according to the procedure described in Section 10.1. Resume
support.
Record the alert message and contact Abbott Medical
representative.
B1
Alert
BATTERY MODULE FAIL
The Console battery will not function. An audible alarm will
sound.
Switch to the backup Console, Motor and Flow Probe
according the procedure described in Section 10.1. Resume
support. Record the alarm message and contact your Abbott