CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 115 of 126
© 2019 Abbott– Document No PL-0047, Rev 11 (December 2019)
Table 16: Console Alarms & Alerts
ID
Alarm/
Alert
Text Message
System Status & Operator Response
M4
Alarm
MOTOR ALARM
An audible alarm will sound and the System will continue to
operate.
Press the alarm acknowledge button, if the visual alarm
message does not disappear, clamp the return tubing, stop the
Pump and switch to the backup Console, Motor and Flow
Probe according the procedure described in Section 10.1.
Resume support. Record the alarm message and contact your
local Abbott Medical representative.
M5
Alert
SET PUMP SPEED NOT
REACHED
Check the Pump flow: If Pump flow is satisfactory; reduce set
speed while insuring that flow is maintained.
Press the alarm acknowledge button. If the alert repeats,
clamp the return tubing, stop the Pump and switch to the
backup Console, Motor and Flow Probe according the
procedure described in Section 10.1. Resume support. Record
the alert message and contact your Abbott Medical
representative.
If the Pump flow is not satisfactory, clamp the return tubing
and switch to the backup Console, Motor and Flow Probe
according to the procedure described in Section 10.1. Resume
support.
Record the alert message and contact Abbott Medical
representative.
B1
Alert
BATTERY MODULE FAIL
The Console battery will not function. An audible alarm will
sound.
Switch to the backup Console, Motor and Flow Probe
according the procedure described in Section 10.1. Resume
support. Record the alarm message and contact your Abbott
Medical representative.
B2
Alert
BATTERY BELOW MINIMUM
The Pump will stop after a very short time.
Plug the Console into a AC power outlet to charge the battery.
If no AC outlet is available, switch to the backup Console,
Motor and Flow Probe according to the procedure described in
Section 10.1.
Resume support.
F1
Alert
FLOW PROBE
DISCONNECTED
Check the Flow Probe connection on back of Console.
If necessary, reconnect the Flow Probe connector to the back
of the Console. Press the alarm acknowledge button. Switch to
the backup Flow Probe, if the alert message repeats.
S3
Alert
SYSTEM ALERT
Press the alarm acknowledge button, if the message does not
disappear, clamp the return tubing, stop the Pump and switch
to the backup Console, Motor and Flow Probe according the
procedure described in Section 10.1. Resume support.
Record the alarm message and contact your local Abbott
Medical representative.