CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 51 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
4. Retain all packaging materials in the event that the Console or any other component
needs to be returned to Abbott Medical for repair or maintenance.
WARNING
Never put containers of liquids on top of or in the immediate vicinity of the
Console. Always prevent liquids from entering the device, since this can cause
permanent damage to the Console.
WARNING
Never operate the System in presence of flammable gases (e.g. flammable
anesthetic mixture), since this could lead to fire and explosions.
CAUTION
Make sure that the System cables (Motor cable, Flow Probe cable, etc.) are
organized properly in order to avoid accidents and reduce the likelihood of EMI.
EMI may interfere with the System and potentially may stop the Motor. Prevent
cable loops on the floor and avoid cables hanging over other equipment or
furniture.
CAUTION
Before each use, verify that the cable connecting the Motor to the Console is not
kinked, which can occur with improper handling such as wrapping the cable
tightly around the Motor. If the cable is kinked, replace the Motor.
WARNING
Make sure that tubing and Cannulas between the System and patient are
secured properly.
WARNING
To avoid the risk of electric shock, do not touch any signal input/output part
(SIP/SOP) like pins and contacts of any electrical connector or socket of this
equipment and the patient simultaneously.