CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 80 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
[400] 2009-08-22 14:16:06 System Started
(a)
(b)
(c)
(d)
Figure 51: Monitor – Sample Data Logger Event
The System records a number of parameters for each event.
a) There are 9 different types of events and each has its own associated icon. The icons
and the associated events are shown in the table below (
).
Table 9: Event Types for the System data logger
Name
Symbol
Event
Uptime
This event is generated every 15 min to give the
total amount of time that the Console has been
powered up.
Info
This event is generated when the System is started,
stopped, turned off and for System time changes.
Alarm
This event is generated when a System alarm is
activated. See Table
13 for a list of alarms.
Alarm acknowledged
This event is generated when the “Alarm
Acknowledge” button is pressed and an alarm is
still active.
Alarm Deactivated
This event is generated when an alarm is no longer
active (i.e. the condition is resolved).
Alert
This event is generated when an alert is activated.
See Table 13 for a list of alerts.
Alert acknowledged
This event is generated when the “Alarm
Acknowledge” button is pressed and an alert is still
active.