Circulatory Support System Operation Manual (US)
Page 104 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
3) Reconnect AC power when cleaning is completed.
WARNING
The Console should not be covered with plastic or insulating material during
use or AC powered storage as it may overheat and malfunction.
9.3 Recommended Preventive Maintenance
The services listed in
are applicable to the main and backup Consoles and
should be performed by qualified personnel trained by Abbott Medical. These
maintenance processes are only to be performed off-patient.
WARNING
Perform routine battery maintenance by following the preventive maintenance
schedule in order to confirm proper calibration of the “battery charge
remaining” indicator.
WARNING
During routine maintenance, verify that the fan is not blocked. A faulty or
blocked fan within the Console may cause overheating, System malfunction, or
trigger an alarm.
Table 15: Console Maintenance Schedule
Required Action
After
Each Use
Every 6
Months
Every 12
Months
Every 2
Years
Perform Battery Maintenance
provided
in
Section 9.4
X
Clean all external surfaces and verify
the general condition of the Console. If
any damage is present return the
Console back to Abbott Medical for
service.
X
X
Verify that all labels on the Console
are present and legible.
X
X
Verify that the leakage currents comply
with the requirements of IEC 60601-1.
Refer to Appendix II
for specific electrical
safety requirements for the Console.
X
Verify that the ground resistance
complies with the requirements of IEC
60601-1.
X
Replace Internal Rechargeable Battery
Pack.
X