24
User Guide
Setting channel level alarms and warnings
Each input channel supports two user-configurable thresholds for both alarms and warnings. It is
your responsibility to define these thresholds and enter the appropriate values in the
Level validation
section.
When entering values for these thresholds, you must ensure that warning values, if ever attained, are
attained before alarm values.
The following logical sequence should be applied:
Minimum alarm value
<
Minimum warning value
<
Maximum warning value
<
Maximum alarm value
.
—
Figure 22
Entering alarm and warning level values
Click
Apply
at the bottom of the page to save your changes.
NOTICE
Only when a value is above its RoC trigger level will a rate-of-change warning or alarm
condition be raised.