4-6
Running a Sample
4.
If blank results are unacceptable for certain parameters, the ana-
lyzer will rerun the blank. If the results are still unacceptable, the
message shown at right appears.
If this occurs, press
OK
, find the warning flag displayed at the
upper right on the screen, then go to the section of
for that flag.
CAUTION:
Do not accept an unacceptable blank (with high values).
Note:
A new blank will be required after the following procedures:
Reagent Pack Change
Soak Cleaning (
see “Soak Cleaning” on page 6-6
)
Deep Cleaning (
see “Deep Cleaning” on page 9-18
)
Self-test
HM5 Reboot