Troubleshooting
9-7
L
Insufficient Lyse reagent
delivered to burst all
RBCs (or lyse-resistant
RBCs), platelet clumping
or large platelets for a
Cavalier King Charles
Spaniel.
• If patient is a Cavalier King Charles Spaniel or other breed
with known predisposition to macro-platelets, perform a
manual smear.
• If the draw was difficult, the platelets are low, or there were
delays in filling the tube, redraw and rerun.
• Make sure the Lyse tube (yellow) has no kinks, or bends and
that nothing is compressing the tubing.
• Check for bubbles in the Lyse tube. If found, prime the Lyse:
see
.
• If no Lyse issues are found, rerun the sample with increased
Lyse +0.1 ml, then if needed with +0.2 ml: see
.
• If the L Flag still persists, draw a new sample, aiming for a
clean needle puncture and fast transfer to the EDTA tube.
• If the
L
flag still persists, call Abaxis Technical Support.
W
Severe platelet clumping
or very large WBCs.
• If the cat platelet his-
togram resembles the
diagram at right,
redraw the sample,
reduce delays in fill-
ing the tube, and mix
well with at least 10-15 inversions immediately.
• If the
W
flag persists, call Abaxis Technical Support.
Y
• Reagent tubing con-
nected to incorrect
port or bottle, clogging,
or insufficient Lyse 2
delivered.
• If the
Y
flag is accom-
panied by many other
warnings (
c
,
C
,
E
,
B
,
Y
), call Abaxis Techni-
cal Support.
• Make sure the Lyse 2 tube (orange) is connected to the
Lyse 2 bottle. If not, disconnect the incorrectly connected
tubes from the analyzer, and hold them up so the liquid in
them flows back into the bottles. Connect tubes to the cor-
rect connections on the analyzer, bottle caps or dip tubes by
color-matching the tubes, analyzer port, and bottle caps or
dip tube heads, then prime Lyse 2.
• If the tubing is already correctly connected, check the Lyse 2
tubing connections at the analyzer and the bottle cap/dip
head. Make sure the tube is securely connected to the con-
nector. If this reagent pack was newly installed, prime the
Lyse 2 twice, then rerun the sample.
• If the Lyse 2 tube is securely connected, run 1-3 automatic
self-cleanings (see
“Automatic Self-Cleaning” on page 6-8
),
then rerun the sample or blank that gave the
Y
warning.
• If the
Y
flag persists, call Abaxis Technical Support.
Z
Lyse 2 delivery issues, or
a dirty Wash Head or
aperture.
• Check the Lyse 2 tube as described above for the
Y
flag.
• If the Lyse 2 tube connected correctly, inspect the Wash
Head and clean as needed (see
), then run 1-3 automatic self-cleanings (see
“Automatic Self-Cleaning” on page 6-8
). Rerun the sample.
• If the
Z
flag persists, call Abaxis Technical Support.
Flag
Meaning
Description / Recommended action
Summary of Contents for VetScan HM5
Page 1: ...Operator s Manual ...
Page 2: ......
Page 4: ...1 2 ...
Page 6: ...1 4 ...
Page 26: ...2 10 Installing the VetScan HM5 ...
Page 38: ...3 12 Settings ...
Page 50: ...4 12 Running a Sample ...
Page 64: ...5 14 Quality Control and Calibration ...
Page 80: ...6 16 Maintenance Service ...
Page 88: ...8 4 Shutdown ...
Page 112: ...9 24 Troubleshooting ...
Page 126: ...A 10 Introduction to Veterinary Hematology ...
Page 136: ...B 10 Operating Principles ...
Page 154: ...D 16 Veterinary Case Studies ...
Page 163: ......