Shutdown
8-3
8.1.2
Longer Term Shutdown or Storage
If the analyzer will be unused for more than two weeks, or will be shipped, use the
Prepare for Ship-
ment
option as follows.
Required materials:
Distilled water, 200 ml
1-liter flask, polypropylene (from Abaxis) or an equivalent container
1.
From the Home screen, select
Exit
.
2.
Select
Prepare for Shipment
.
3.
Follow the instructions that appear on the display.
The procedure drains the reagents from the analyzer, and flushes the Diluent from the internal
tubing. An alarm sounds when it is safe to turn off the analyzer using the power switch on the
back panel.
8.1.3
Powering Off for Troubleshooting
On occasion, the analyzer may freeze or require a reboot for troubleshooting. In these situations, pro-
ceed as follows:
1.
Turn the analyzer off using the power switch on its back panel.
2.
Turn the analyzer on.
CAUTION:
If this is not done before the analyzer is powered off for long peri-
ods, salt crystals can form blockages in the tubing.