10-4
Specifications
10.3 Precision
The following precision parameters were established using normal level control on one instrument,
with ten replicate measurements performed in one day.
Table 10-3: Control Parameter Precision
Parameter
Mean
SD
%CV
WBC (10
3
cell/
μ
l)
8.05
0.15
1.89%
RBC (10
6
cells/
μ
l)
4.49
0.111
2.44%
EOS (10
3
cell/
μ
l)
5.00
0.07
1.32%
HCT (%)
40.39
1.01
2.51%
HGB (g/L)
125.66
1.71
1.36%
MCV (fL)
90.03
0.50
0.56%
PLT (10
3
cell/
μ
l)
248.36
18.30
7.37%
MPV (fL)
12.69
0.177
1.39%