Maintenance & Service
6-3
6.1.1
Scheduling Maintenance
The analyzer’s maintenance reminder intervals are preset, but can be adjusted to best suit the particular
needs of the clinic.
1.
From the Home screen, select
Settings > Admin Settings
.
2.
Selecting the
Scheduling
field changes the setting from
Enabled
to
Disabled
and vice versa. Set this to
Enabled
.
3.
Select
Schedule Settings
.
4.
As needed, select the appropriate fields to set the
Interval
and
Cleaning Day and Time
for the Wash Head and Soak Cleanings.
Abaxis recommends these intervals:
Cleaning the Wash Head: 7 days
Soak Cleaning: 14 days
5.
Select
Accept
for these settings then
Accept
again to confirm the Scheduling choice.
Note:
If your sample run rate will be less than
5 samples per day, you may set the Soak
Cleaning Interval to 30 days.
Note:
On installation, you can reset both timers to begin from the current
date by pressing
Reset
. The next Scheduled Maintenance will then
be a full interval away from the current date.
Summary of Contents for VetScan HM5
Page 1: ...Operator s Manual ...
Page 2: ......
Page 4: ...1 2 ...
Page 6: ...1 4 ...
Page 26: ...2 10 Installing the VetScan HM5 ...
Page 38: ...3 12 Settings ...
Page 50: ...4 12 Running a Sample ...
Page 64: ...5 14 Quality Control and Calibration ...
Page 80: ...6 16 Maintenance Service ...
Page 88: ...8 4 Shutdown ...
Page 112: ...9 24 Troubleshooting ...
Page 126: ...A 10 Introduction to Veterinary Hematology ...
Page 136: ...B 10 Operating Principles ...
Page 154: ...D 16 Veterinary Case Studies ...
Page 163: ......