Running a Sample
4-5
4.4 Running a Blank Measurement
A blank measurement must be run at the beginning of the day before any samples can be run, or every
12 hours for 24-hour clinics. Blanks must also be run after a reagent pack change. The blank checks the
cleanliness of the VetScan HM5’s fluidic system, and establishes a baseline for sample measurements.
The results of a blank are used to determine if the background will affect the test results, and whether
the analyzer needs cleaning or maintenance.
1.
From the Home screen, select
Measure > Run
.
2.
If the
Blank results are expired
s
creen appears, select
OK
to run a blank.
(If a blank has already been run and accepted for the day, this message does not appear.
To run a new blank, select
Run > Blank
.)
3.
When the blank has run, if the blank is successful (meaning there are no values reported in red
with
+
or
*
signs, and no warning message displayed) the results will appear on the screen.
Note:
Each blank measurement is valid for 12 hours, after which the ana-
lyzer displays
Blank results are expired!
Accept the running of the
blank then accept the blank to begin running samples again.
In addition, a new blank must be run each time the analyzer is powered
off and then powered on again.
Summary of Contents for VetScan HM5
Page 1: ...Operator s Manual ...
Page 2: ......
Page 4: ...1 2 ...
Page 6: ...1 4 ...
Page 26: ...2 10 Installing the VetScan HM5 ...
Page 38: ...3 12 Settings ...
Page 50: ...4 12 Running a Sample ...
Page 64: ...5 14 Quality Control and Calibration ...
Page 80: ...6 16 Maintenance Service ...
Page 88: ...8 4 Shutdown ...
Page 112: ...9 24 Troubleshooting ...
Page 126: ...A 10 Introduction to Veterinary Hematology ...
Page 136: ...B 10 Operating Principles ...
Page 154: ...D 16 Veterinary Case Studies ...
Page 163: ......