Pannoramic 250 FLASH II 1.18.1 User’s Guide
4 Software Description
To insert and remove slides
•
In Automatic mode a preloaded magazine has to be placed into the main unit. This magazine
is the first one, irrespectively of the total number of magazines (1-10) to be loaded.
If a magazine is present in the input feed, the
button turns active. Click this button to
load the magazine.
•
If you do not want to load all slides, you can specify the first and last slide to be loaded in the
Scan Range
by entering the appropriate slide numbers in the selection boxes of
First slide
and
Last slide
.
•
To move all magazines to the output stack, click
. You can save time if you
remove all magazines from the input stack before you click this button.
You also find the
button on this panel, that you can start scanning with, but it is
recommended to go through all the settings before clicking it.
To administer profiles, use the
Scan Profiles
field.
•
The current profile name is displayed at the bottom left corner of the screen.
•
To save the current profile, click
Save
.
•
To select an existing profile from the selection menu, click
Load
.
•
You can delete the current profile by clicking
Delete
.
•
To access the Scan Options window, click
.
Note:
These settings are only applicable if you do not use profiles. If you use
profiles, the settings stored in them are used. The used profile name is
indicated at the bottom left of the main window.
For more information, see section
Filter Settings
•
Focus channel –
Choose a filter that you want to focus with. It is recommended to choose a
filter that is related to the staining that stained the most of the specimen, or where the
exposition time is low. You can choose from filters that you have previously selected in the
Channel Settings
field.