Pannoramic 250 FLASH II 1.18.1 User’s Guide
4 Software Description
4.3.3 Preview panel
The Preview panel offers options for threshold setting and selection of the area to be scanned.
Figure 35 – Preview panel
Use threshold
•
Brightfield preview –
a normal type of illumination in the microscope for scanning the slide.
•
Darkfield preview –
the preview image is created with transmitted Darkfield illumination.
This option is recommended for you to choose in case the sample is not correctly stained (for
example, the specimen is lacking contrast) when using Brightfield illumination. The image is
created within a minute.
•
Scan specimen with auto threshold
–
This automatically detects the specimen in
the preview image.
Important!
Make sure that the result of detection is acceptable for you. Otherwise set
the threshold value manually or append missing areas with the help of
Scanned
function located at the preview toolbar.
May 29, 2014 – Rev. 1
3DHISTECH Ltd.
57
(
137
)