Pannoramic 250 FLASH II 1.18.1 User’s Guide
5 Troubleshooting and Maintenance
•
Focus range setting is incorrect
•
Change focus range setting.
All individual images show a shading effect
•
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide (see section
).
5.2.1 Fluorescent Applications (optional)
The switching of the Lumencor cannot be triggered, in spite of trigger connection through the
camera.
•
Contact 3DHISTECH Service for support.
No suitable exposure time can be found. Signal is too weak.
•
HXP 120: Shutter is closed
•
The opening of the iris contained in the lamp module for fluorescent applications
can be varied by means of the wheel located at the front side of the module. Open
the iris completely.
•
HXP 120: Intensity is too low
•
Increase intensity with the rotary button at the front of the HXP 120.
•
HXP 120 is switched off
•
Switch light source on.
•
When using Lumencor
•
Contact 3DHISTECH Service for support.
May 29, 2014 – Rev. 1
3DHISTECH Ltd.
120
(