Pannoramic 250 FLASH II 1.18.1 User’s Guide
4 Software Description
Filter name
The name of the filter is displayed here.
Exposure Settings
Exposure time
–
You can set the fix exposition time here. By selecting the
Auto
button, the exposition time is set to
<AUTO>
(the exposition time is
calculated by the software). If the fix position is set, the
Exposure time
weight
value is not valid.
You can set the
Exposure time weight (%)
here (displayed under
TW
in the
Filter Selection
frame).
In full automatic mode, the system automatically calculates the exposition
time to the given channel. You can increase / decrease this amount here in a
range of 50%-200%.
Multilayer Settings
Enable Multilayer
function and the concerning settings can be activated in
this field, accordingly, you can set the
Extended focus
and
Z-stack
values for
each channel separately.
Digital Gain
Note:
This function is camera-dependent, works only with
AxioCam MRm Rev 3 camera (Index value ranges from 0
to 4) and pco.edge 4.2 (Index values: 0-6).
You can enter the
Index
value by entering it in the box or by selecting the
value with the arrow buttons. If you raise the
Index
value, the exposition
time is halved by every step, and the image becomes noisier.
You can move between the filters by clicking
Previous Filter
or
Next Filter.
May 29, 2014 – Rev. 1