5 Troubleshooting and Maintenance
5 Troubleshooting and Maintenance
5.1 Safety Information
Check cables, plugs and connections before operation, and if defective or cannot be replaced, contact
3DHISTECH Service for support.
Check whether the device is compatible with your local line voltage. It can operate at line voltages
100V to 240V and 50/60 Hz.
Only plug the device in a socket with protective grounding (earth) contact.
Do not operate the device in explosive environment.
Disconnect the device from the power outlet before opening it or replacing a part.
The product is not equipped with protection against hazardous materials (infectious, toxic,
radioactive, corrosive substances or other effects damaging to health). Decontaminate infectious
materials after use.
Dust, specks and humidity may decrease or harm the operating capability of the device. Protect it
with dust cover when not in use. Switch the power off before covering it.
There is crush hazard during operation in the area of input or output. Inserts/magazines should be
moved through software. The protective shield of the magazine feed unit can be unmounted when
slide scanning has stopped. The device can be opened for service reasons only when switched off,
and in cases stated in this User’s Guide (for example, changing filters).
Defective devices do not classify as domestic waste. Dispose of the device in compliance with the
currently valid legal requirements.
Lumencor illumination unit
•
NEVER
switch on the Lumencor SPECTRA light source when it is not mounted to the
microscope.
•
Do not cover the unit, as it would cause abnormal temperature rising.
•
Ensure that the light source is mounted well and the fixing lever is closed.
HXP 120 illumination unit
•
Keep inflammable and combustible materials outside the range of HXP 120 illumination unit.
•
Do not expose yourself to the UV radiation of the HXP 120 illumination unit.