Figure 94 – Removing a slide
4. Press clamping lever
(4)
down. The slide position lock releases.
5. Retrieve slide
(3)
from the holder.
6. Release clamping lever
(4)
.
7. Turn down the slide loader
(2)
back into the original position and pull it inwardly (to the
Pannoramic 250 FLASH II), until the guiding pin connects with the motor block groove again.
8. Press the slide loader
(2)
down to mechanical stop position, in which the locking lever must
engage
(1)
. Use moderate force to press the loader part opposite the motor in clamp
direction, until this part mechanically engages, too.
9. Close the protective cover again (see section
5.3.1 Opening the protective cover
).
10. Launch the control software.