Pannoramic 250 FLASH II 1.18.1 User’s Guide
5 Troubleshooting and Maintenance
5.3.5 Installing and Removing the Filter Module (Fluorescent Option)
Figure 95 – Installing and removing the filter module
1. If using an HXP 120 illumination unit, turn it off.
Warning!
Crush hazard! Stop the running digitization process before you proceed
with filter module installation or removal.
2. Open the protective cover (see section
5.3.1 Opening the protective cover
).
3. Select the desired filter in the corresponding software window (see section
Microscope Settings/Filter Settings panel
). The filter will be positioned in such a way that
the user can access it without having to turn the filter wheel manually.
4. Enter the respective parameters in the input fields.
5. Hold the filter module
(3)
by its handling pads (one on the right, one on the left (lateral)
side of the module) and insert it in forward tilted position from above into the upper spring-
loaded clamps
(1)
at the filter wheel.
Warning!
Avoid touching the surface of the filters, and store the modules in their
own boxes when not used.