5 Troubleshooting and Maintenance
6. Press the filter module down, until it safely connects to the bottom spring-loaded clamps
(2)
of the filter wheel. Mounting position of the filters: excitation filter
(5)
and emission
filter
(4)
7. To remove the filter module, first slightly tilt it out of its upper spring-loaded fixtures, then
pull it out of the bottom spring-loaded components and take it off.
8. Close the protective cover (see section
5.3.1 Opening the protective cover
).
9. If an HXP 120 illumination unit is used, turn it on again (consider the cooling time possibly
required).
5.3.6 Removing a Magazine from the Transporting System
Figure 96 – Removing a magazine
1. Open the protective cover (see section
5.3.1 Opening the protective cover
).
Warning!
Crush hazard! Remember to stop the running digitization
process,
terminate the software session and turn power supply off before you
remove a jammed magazine (see section
).
Caution!
Check that the magazine is not tilted, so that it rests on both horizontal
guide rails.
May 29, 2014 – Rev. 1
3DHISTECH Ltd.
128
(