Pannoramic 250 FLASH II 1.18.1 User’s Guide
4 Software Description
•
Stitching channel –
This is done on one channel. You can choose whether you want to do
stitching on the focus channel, or on another channel. It is recommended to choose
another channel if the focus channel was determined on the basis of exposition time, and it
would display less information. It is recommended to choose a filter that is related to the
staining that stained the most of the specimen.
The default selection is
<
Focus channel>
.
Objective position
Set the objective changer to the appropriate position in this field.
Channel Settings
Select what types of filters you want to use during scanning. You can select the filter types
and the number of channels to be used.
Those filters are displayed automatically that you have previously defined in the
Options
/
Microscope Settings
at the mode selection screen.
For more information on predefining filters, see section
4.2.3 Microscope Settings / Filter
To set filter properties, select a filter and click on any of the white boxes of the ET, TW and
DG columns next to the filter name. This displays the Filter Properties window.
Figure 69 – Filter properties window
May 29, 2014 – Rev. 1
3DHISTECH Ltd.