April 2019
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
24-5
Technical Data
Active Noise Suppression
RL drive (< 5 µA)
Pulse Tone
Yes
Respiration
Method
Impedance Pneumography
Input Connector
Same as ECG
Sensing Lead
II
RR Resolution
1 bpm (breaths per minute)
Measurement Range
0 to 120 bpm
Measurement Accuracy
±3 bpm
Measurement Sensitivity
0.25 ohms (minimum)
Report Interval
1 second
Bandwidth
0.17 to 3.3 Hz (-3dB)
Impedance Measuring Current
40 μA @ 28 kHz square wave across Lead II
Pulse Oximetry
Method
Absorption
– Spectrophotometric (dual wavelength)
(Functional oxygen saturation of arterial hemoglobin)
Input Connector
9-pin connector
SpO
2
/ PR Resolution
SpO
2
: 1 O
2
%
PR: 1 bpm (beat per minute)
Measurement Range
SpO
2
: 20 to 100%
PR: 30 to 240 bpm
Measurement Accuracy
SpO
2
: from 70 to 100%: ±2% (O
2
%), < 70%: unspecified
PR: ± 3 bpm
Measurement Test Method
Comparison versus co-oximeter, per ISO 80601-2-61
Report Interval
1 second. Numeric values held < 30 seconds
Pulse Tone
Yes (pulse tone pitch tied to SpO
2
parameter value)
Alarm Signal Generation Delay
< 0.5 sec
Non-Invasive Blood Pressure
Method
Oscillometric
Input Connector
Single Lumen Hose (Quick-Disconnect fitting)
Содержание PPM3
Страница 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Страница 4: ......
Страница 16: ......
Страница 28: ......
Страница 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Страница 52: ......
Страница 96: ......
Страница 100: ......
Страница 123: ...April 2019 Nightingale PPM3 User s Guide 11 23 Working With Menus Figure 48 Setup Alarms Menu Screen 2 ...
Страница 124: ...11 24 Nightingale PPM3 User s Guide April 2019 Working With Menus Figure 49 Setup Alarms Menu Screen 3 ...
Страница 138: ......
Страница 160: ......
Страница 164: ......
Страница 184: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......