11-10
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
Working With Menus
11.1.6. Masimo CO
2
/ Capnography
Figure 37. Setup CO
2
Menu With Masimo Capnography
Setting
Description
Lower Limit
ETCO
2
& RRc alarm lower limits
Upper Limit
ETCO
2
, RRc, & FICO
2
alarm upper limits
Auto
Auto-adjust lower limits to 80% of current values and upper
limits to 125% of current values.
Alarms On
Enables (Yes) or disables (No) alarms
Print on Alarm
Print a strip chart summary upon alarm conditions
Size
Changes display amplitude of CO
2
waveform
Apnea Time
Time since last breath before RRc alarm annunciates
O
2
Compensation
Adjust CO
2
value based on this range O
2
% in fresh gas
N
2
O Compensation
Adjust CO
2
value based on this range N
2
O% in fresh gas
Zero Set
Initiate manual zeroing of CO
2
sensor (see Masimo CO
2
Содержание PPM3
Страница 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Страница 4: ......
Страница 16: ......
Страница 28: ......
Страница 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Страница 52: ......
Страница 96: ......
Страница 100: ......
Страница 123: ...April 2019 Nightingale PPM3 User s Guide 11 23 Working With Menus Figure 48 Setup Alarms Menu Screen 2 ...
Страница 124: ...11 24 Nightingale PPM3 User s Guide April 2019 Working With Menus Figure 49 Setup Alarms Menu Screen 3 ...
Страница 138: ......
Страница 160: ......
Страница 164: ......
Страница 184: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......