April 2019
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
11-3
Working With Menus
HR Source
Calculate HR/PR from ECG (default
‒ green), SpO
2
(cyan), or ART-
labeled IBP (red). See Figure 28 for HR/PR parameter box
appearance when HR Source is SpO
2
or ART.
Pulse Tone Source
Enables audible tones for each detected beat. When source is
SpO
2
, the tone pitch varies (higher SpO
2
% = higher pitch).
ECG Filter
Set to either Monitor (0.67
– 40 Hz) or Diagnostic (0.05 – 40 Hz)
Cal
Injects a simulated 1 mv calibration pulse into the ECG waveform
Figure 32. HR/PR Values Sourced From SpO2 or ART