24-10
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
Technical Data
IBP Resolution
1 mmHg
Pulse Rate Measurement Range
30 - 250 bpm
Pulse Rate Accuracy
± 2 bpm or ±2 %, whichever greater
Numeric Update Rate
Every 3 seconds
Temperature
Compatibility
YSI 400-series probes
Measurement Mode
Direct (as defined in ISO 80601-2-56)
Input Connector
2-pin connector
Display Units
°F and °C (user-selectable)
Measurement Resolution
0.1 °F (0.1 °C)
Measurement Range
41.0 to 122.0 °F (5.0 to 50.0 °C)
Measurement Accuracy
±0.2 °F (±0.1 °C) plus probe tolerance
Transient Response
Within 30 seconds from 25 to 27 °C