11.1.4. Non-Invasive Blood Pressure
Figure 35. Setup NBP Menu
Setting
Description
Lower Limit
NBP systolic/mean/ diastolic alarm lower limits
Upper Limit
NBP systolic/mean/ diastolic alarm upper limits
Auto
Auto-adjust NBP lower limits to 80% of current values and upper
limits to 125% of current values.
Alarms On
Enables (Yes) or disables (No) alarms
Print on Alarm
Print a strip chart summary upon alarm conditions