April 2019
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
14-1
Viewing Trends
14. Viewing Trends
Each time the PPM3 takes a NBP blood pressure measurement or at user-
selectable intervals, it stores the following physiological parameters (provided
they are in use or are enabled) in the trends table along with the current time:
•
Heart Rate or Pulse Rate
•
Non-Invasive Blood Pressure (s/d/m)
•
Respiration Rate
•
Oxygen Saturation (SpO
2
)
•
End-tidal CO
2
(ETCO
2
)
•
Fractional inspired CO
2
(FICO
2
)
•
Integrated Pulmonary Index (IPI)
•
Temperature (T1)
•
Invasive Blood Pressure (P1 s/d/m and P2 s/d/m)
If a HR value is being sourced from ECG, the trends table will show a HR column
labeled in green.
If PR is being sourced from SpO
2
, there will be a PR column labeled in cyan.
If PR is being sourced from an ART-labeled IBP, there will be a PR column
labeled in red.
Up to 72 hours of trends data may be stored on the PPM3. All trend samples can
be manually deleted via the trends menu. They are also deleted when a patient is
discharged at the MPC.
Содержание PPM3
Страница 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Страница 4: ......
Страница 16: ......
Страница 28: ......
Страница 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Страница 52: ......
Страница 96: ......
Страница 100: ......
Страница 123: ...April 2019 Nightingale PPM3 User s Guide 11 23 Working With Menus Figure 48 Setup Alarms Menu Screen 2 ...
Страница 124: ...11 24 Nightingale PPM3 User s Guide April 2019 Working With Menus Figure 49 Setup Alarms Menu Screen 3 ...
Страница 138: ......
Страница 160: ......
Страница 164: ......
Страница 184: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......