3-8
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
ECG and Heart Rate Monitoring
3.7. Electrode Placement for Paced Patients (IEC)
For ECG cable lead sets with IEC (Europe) lead designations, the optimal
electrode placement for patients with pacemakers may be as illustrated in the
following diagram:
Note
– If using a three-wire ECG cable lead set, you only need to apply the R, L, and F
electrodes, and only Lead II will be available.
Figure 18. Paced IEC Electrode Placement
1.
Position the right arm (R) electrode on the right mid-clavicular line, 5
th
intercostal space.
2.
Position the left arm (L) electrode on the left mid-clavicular line, directly
below the clavicle.
3.
Position the neutral (N) electrode on the right mid-clavicular line, 6
th
and
7
th
intercostal space (5-lead cable only).
4.
Position the left leg (F) electrode on the left mid-clavicular line, 6
th
and 5
th
intercostal space.
5.
Position the chest (C) electrode on the 4
th
intercostal space, right sternal
border (5-lead cable only).
N (Black)
R (Red)
L (Yellow)
F (Green)
C (White)
Содержание PPM3
Страница 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Страница 4: ......
Страница 16: ......
Страница 28: ......
Страница 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Страница 52: ......
Страница 96: ......
Страница 100: ......
Страница 123: ...April 2019 Nightingale PPM3 User s Guide 11 23 Working With Menus Figure 48 Setup Alarms Menu Screen 2 ...
Страница 124: ...11 24 Nightingale PPM3 User s Guide April 2019 Working With Menus Figure 49 Setup Alarms Menu Screen 3 ...
Страница 138: ......
Страница 160: ......
Страница 164: ......
Страница 184: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......