8-12
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019
Masimo CO
2
Monitoring
Message
Parameter
Value
Possible Cause
Suggested Action
CO
2
Occluded Line
---
ISA CO
2
sampling line
cannot be cleared due to
moisture or other
obstruction.
Replace the sampling line. If connected
to scavenging system, disconnect to see
if message disappears.
CO
2
No Sampling Line
---
ISA CO
2
sampling line
has become unplugged
from the LEGI connector
on the front of the
module.
Reconnect the sampling line.
CO
2
Check Adapter
---
1) The lens within IRMA
airway adapter has
become fogged or
blocked.
2) IRMA airway adapter
has become unplugged
from the sensor.
Replace or reconnect the airway adapter.
CO
2
Needs Zeroing
---
CO
2
zero offset needs to
be removed.
Follow the zeroing procedure provided in
this chapter.
CO
2
Check Sensor
---
ISA module or IRMA
sensor have reported an
internal error.
Disconnect the module or adapter from
the PPM3 and then reconnect. Contact
technical support if the problem
continues.
CO
2
problem detected
---
CO
2
interface has
encountered a problem.
Check the CO
2
subsystem including
sampling line and exhaust port. Verify
that the exhaust port is not blocked.
Power-cycle the PPM3. Contact
technical support if the problem
continues.
Содержание PPM3
Страница 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Страница 4: ......
Страница 16: ......
Страница 28: ......
Страница 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Страница 52: ......
Страница 96: ......
Страница 100: ......
Страница 123: ...April 2019 Nightingale PPM3 User s Guide 11 23 Working With Menus Figure 48 Setup Alarms Menu Screen 2 ...
Страница 124: ...11 24 Nightingale PPM3 User s Guide April 2019 Working With Menus Figure 49 Setup Alarms Menu Screen 3 ...
Страница 138: ......
Страница 160: ......
Страница 164: ......
Страница 184: ......
Страница 194: ......
Страница 196: ......
Страница 198: ......