25
The reference figures are present on pages 3, 4, 5 and 6 of this manual and in the product Quick Guide.
EN
hurrimix
2
hurrimix
2
115Vac/60hz
Manufacturer’s declaration and guidance – electromagnetic immunity
Hurrimix
2
is designed to be used in the electromagnetic environment specified below. The client or user must ensure that it is used in such an environment.
IMMUNITY
test
IEC 60601-2 level
Compliance
level
Electromagnetic Environment - Guidance
Conducted
RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
V
1
= 3 V rms
E
1
= 3 V / m
Portable and mobile RF communications equipment should be used no closer to any of Hurrimix
2
’s
parts, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = [3,5 / V
1
] √P
d = [3,5 / E
1
] √P from 80 MHz to 800MHz
d = [7/ E
1
] √P from 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in Watts (W), according to the
transmitter manufacturer, and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
(a)
,
should be less than the compliance level in each frequency range
(b)
.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and
people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which Hurrimix
2
is used exceeds the applicable RF
compliance level above, the product’s normal operation should be observed. Additional measures, such as reorienting or relocating the product, may be
required.
b) The field strength over a frequency ranging from 150 kHz to 80 MHz should be less than 3 V/m.
hurrimix
2
230Vac/50hz
Manufacturer’s declaration and guidance – electromagnetic immunity
Hurrimix
2
is designed to be used in the electromagnetic environment specified below. The client or user must ensure that it is used in such an environment.
IMMUNITY
test
EN 60601-2 level
Compliance
level
Electromagnetic Environment - Guidance
Conducted
RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
V
1
= 3 V rms
E
1
= 3 V / m
Portable and mobile RF communications equipment should be used no closer to any of Hurrimix
2
’s
parts, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = [3,5 / V
1
] √P
d = [3,5 / E
1
] √P from 80 MHz to 800MHz
d = [7/ E
1
] √P from 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in Watts (W), according to the
transmitter manufacturer, and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
(a)
,
should be less than the compliance level in each frequency range
(b)
.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and
people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which Hurrimix
2
is used exceeds the applicable RF
compliance level above, the product’s normal operation should be observed. Additional measures, such as reorienting or relocating the product, may be
required.
b) The field strength over a frequency ranging from 150 kHz to 80 MHz should be less than 3 V/m.